Appendix
Operating manual AVERO Phoenix
V2.1 E
Page 32 of 35
Recommended distance between portable and mobile communications equipment and
the AVERO Phoenix (Table 206)
The AVERO Phoenix is intended for use in an electromagnetic environment with controlled HF
interferences. The customer or the user of the AVERO Phoenix can avoid electromagnetic
interference by respecting and observing the minimum distance between portable and mobile
HF telecommunication devices (transmitters) and the AVERO Phoenix depending on the rated
output of the communication device as given below.
Rated output
of the transmitter
W
Separation distance depending on the transmitting frequency
in m
150 kHz to 800 MHz
d=0.35√P
80 MHz to 800 MHz
d=1.2√P
800 MHz to
2.5 GHz
d=2.3√P
0.01
0.1
1
10
100
0.04
0.11
0.35
1.1
3.5
0.12
0.38
1.2
3.8
12
0.23
0.73
2.3
7.3
23
For transmitters rated at a maximum output power not listed above, the recommended
separation distance
d
in meters (m) can be determined using the equation applicable to the
frequency of the transmitter, where
P
is the maximum output power rating of the transmitter in
watts (W) according to the specifications given by the transmitter manufacturer.
NOTE 1:
At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
NOTE 2:
These guidelines could not apply to all situations. Electromagnetic propagation
is affected by absorption and reflection from structures,
objects and people.
Содержание AVERO Phoenix
Страница 24: ...Appendix Operating manual AVERO Phoenix V2 1 E Page 24 of 35 13 2 Dimensions of the AVERO Phoenix 190...
Страница 25: ...Appendix Operating manual AVERO Phoenix V2 1 E Page 25 of 35 13 3 Dimensions of the AVERO Phoenix 170...
Страница 27: ...Appendix Operating manual AVERO Phoenix V2 1 E Page 27 of 35 13 5 Declaration of Conformity...
Страница 35: ...BEKA Hospitec GmbH Am R benmorgen 3 35582 Wetzlar Germany Phone Fax 49 0 6 41 9 22 20 0...