PN A51387A
B
3-35
Installing and Calibrating an Aperture Tube
Running a Concentration Control Sample
3
4.
In the Concentration Control window, enter the Lot Number, the Concentration Assay Value,
and the Expiration Date listed on the assay sheet.
5.
Click
(
Start
)
to begin the concentration control run.
6.
The Multisizer 4 software flushes the system and begins the run.
7.
When the run is complete, a second Concentration Control window opens and displays results.
The window displays a message that the concentration is low or high if the Analyzer
measurement deviates from the concentration control by 10% or more. If this occurs, contact
Beckman Coulter Service. For information on how to do this, see
on
page
.
8.
Click
(
Close
)
.
Содержание Multisizer 4
Страница 12: ...xii List of Figures ...
Страница 14: ...xiv List of Tables ...
Страница 22: ...PN A51387AB xxii Introduction Before You Start Warnings and Cautions ...
Страница 42: ...PN A51387AB 1 20 Analyzer Overview Preparing the Analyzer for Sample Runs ...
Страница 90: ...PN A51387AB 3 36 Installing and Calibrating an Aperture Tube Running a Concentration Control Sample ...
Страница 160: ...PN A51387AB 6 16 Analyzing a Sample Using Blank Runs ...
Страница 232: ...PN A51387AB 10 8 Regulatory Compliance Additional Security Features ...