PN A51387A
B
6-8
Analyzing a Sample
Analyzing a Sample
To enter sample specifications:
1.
Select
Sample > Create Sample Specifications
on the Main Menu bar. The Create Sample
Specifications window opens.
2.
In the Create Sample Specifications window:
a.
In the Statistics Type pane, enter the type of bin weighing and the particle size range for
comparison.
NOTE
Settings that you select in the Create Sample Specifications window will not affect or be
affected by selections you make in Preferences page
. Statistics you select in the
Preferences Wizard or Edit Preferences window control view and print settings only.
b.
Enter a title for the sample specifications, if desired.
c.
Select one or more options to specify the sample specification value, and enter a Lower
Limit, Upper Limit, and Nominal Value for the specifications you select.
d.
To save the sample specifications to a file, click
(
Save
)
. Specifications files are identified
with the extension .spec.
e.
Click
(
OK
)
.