Troubleshooting
Resolving Common Problems
Setting Custom Measurement Intervals
If you want to measure gain stage offsets more frequently, you can set custom intervals.
To reset the default measurement intervals:
1.
Select
Run > Troubleshooting > Enter Service Mode
on the Main Menu bar.
NOTE
Entering Service Mode will allow you to change default settings in the Troubleshooting menu.
If you do not enter Service Mode, the Gain Stage Offsets settings will appear as read-only values.
2.
Select
Run > Troubleshooting > Aperture Resistance Settings
on the Main Menu bar. The
Aperture Resistance Settings window opens.
3.
In the Gain Stage Offsets pane, enter new values in the fields provided.
4.
If you are not running Service Mode, a
(
Read-only
)
button appears. If you are running Service
Mode, click
(
OK
)
to save your settings.
System Error: Unable to Measure Gain Stage Offsets
If the Analyzer is unable to measure gain stage offsets automatically, the Multisizer 4 software will
display the following error message:
Unable to measure gain stage offsets
If this error message appears, power off the Analyzer, close the Multisizer 4 software, and restart
the system. If restarting the system does not resolve the problem, contact Beckman Coulter Field
Service.
Содержание Multisizer 4
Страница 12: ...xii List of Figures ...
Страница 14: ...xiv List of Tables ...
Страница 22: ...PN A51387AB xxii Introduction Before You Start Warnings and Cautions ...
Страница 42: ...PN A51387AB 1 20 Analyzer Overview Preparing the Analyzer for Sample Runs ...
Страница 90: ...PN A51387AB 3 36 Installing and Calibrating an Aperture Tube Running a Concentration Control Sample ...
Страница 160: ...PN A51387AB 6 16 Analyzing a Sample Using Blank Runs ...
Страница 232: ...PN A51387AB 10 8 Regulatory Compliance Additional Security Features ...