PN A51387A
B
4-27
Selecting Analysis Settings: SOM and SOP
Using a Standard Operating Procedure (SOP)
4
4.
Click
(
Next
)
to proceed to the next window in the Wizard, Select an SOM File.
5.
In the Select an SOM File window, click
(
Select
)
.
6.
In the Load a Standard Operating Method (SOM) window, navigate to the SOM you want to
include, and click
(
Open
)
. The name of the selected SOM file appears in the Wizard.
7.
Click
(
Next
)
to proceed to the next window in the Wizard, Select a Preference File.
8.
In the Select a Preference File window, click
(
Select
)
.
9.
In the Load Preferences window, navigate to the Preferences file you would like to include, and
click
(
Open
)
. The name of the selected Preferences file appears in the Wizard.
10.
Select Do not allow operator to change preferences to safeguard the format of print and view
settings.