PN A51387A
B
2-8
Software Overview
Status Panel
Main Toolbar Buttons: Fourth Group
The fourth group, at the far right of the Main Toolbar, duplicates functions available in the
Calibration drop-down menu.
NOTE
If you are running the Multisizer 4 software in security mode, certain Main Toolbar buttons will
appear only to user types with sufficient privileges (for example, the
(
Kd Calib.
)
button). The
system administrator sets privileges for each user type.
In the table below, Main Toobar buttons that may not be available based on security settings are
italicized.
Status Panel
The Status Panel, located on the left side of the Multisizer 4 software window, displays elapsed run
time, particle count, and concentration information during an analysis run.
(
Pulses
)
Display electrical pulses as a graph or listing.
(
Zoom
)
Display only the area of the graph between two cursors. Before clicking the
(
Zoom
)
button, first insert two cursors on a graph as follows: a single click
within the graph area creates a single cursor. Click, drag, and un-click to
create two cursors: one at the click point, the other at the un-click point.
After displaying a zoomed image, you can click the
(
Zoom
)
button again
to view the full size distribution, or insert new cursors and click the
(
Zoom
)
button to see a smaller portion of the size distribution.
Table 2.3
Main Toolbar: Buttons (Third Group)
Click...
To...
Table 2.4
Main Toolbar: Buttons (Fourth Group)
Click...
To...
(
Kd Calib.
)
Open the Calibrator Size window and calibrate the aperture.
(
Kd Verify
)
Open the Calibrator Size window and verify aperture calibration.
(
CC Conc.
)
Open the Concentration Control window and run a concentration control
sample.
Содержание Multisizer 4
Страница 12: ...xii List of Figures ...
Страница 14: ...xiv List of Tables ...
Страница 22: ...PN A51387AB xxii Introduction Before You Start Warnings and Cautions ...
Страница 42: ...PN A51387AB 1 20 Analyzer Overview Preparing the Analyzer for Sample Runs ...
Страница 90: ...PN A51387AB 3 36 Installing and Calibrating an Aperture Tube Running a Concentration Control Sample ...
Страница 160: ...PN A51387AB 6 16 Analyzing a Sample Using Blank Runs ...
Страница 232: ...PN A51387AB 10 8 Regulatory Compliance Additional Security Features ...