3-6
Operation
Action
Result
1. Press [
TEMP
].
The last digit of the
TEMP °C
display flashes.
2. Press [
S
] or [
T
]
until
the required temperature
appears, then press [
ENTER
].
Entered temperature appears on the
TEMP °C
display.
SETTING ACCELERATION AND DECELERATION RATES
You can select one of ten acceleration rates and ten deceleration rates
to protect the gradient and sample-to-gradient interface. Select rates
depending on the type of run you are performing—for pelleting runs,
where sample mixing is not a concern, maximum deceleration can be
used. If running delicate gradients, a lower setting may be needed.
Typical acceleration and deceleration rates are listed in Table 3-1. If
no rate is selected, the centrifuge automatically accelerates and decel-
erates at maximum.
Setting Acceleration Rate
Action
Result
1. Press [
ACCEL
]. The
ACC/DEC
display flashes.
2. Press [
S
] or [
T
]
until
the required acceleration
rate appears, then press
[
ENTER
].
Entered number appears on the
ACC/DEC
display.