3-4
Operation
NOTE
➠
If you select the TS-5.1-500 rotor (selection
code S-5.1) a submenu will appear. Select BUC5
for buckets or PLA for microplate carriers.
Action
Result
3. When the selected rotor
number is displayed,
press [
ENTER
].
Rotor number will be shown in the
SPEED
display. You can set
parameters for the run.
SETTING RUN SPEED
You can set centrifuge speed, in either rpm or RCF, for up to the
maximum rated speed of the selected rotor.
Setting RPM
Action
Result
1. Press [
RPM
].
The last digit on the
SPEED
display flashes.
2. Press [
S
] or [
T
]
until
the required speed appears,
then press [
ENTER
].
During centrifugation, the
SPEED
display shows the actual run speed
of the rotor in rpm. The corresponding RCF will be automatically
calculated by the centrifuge.
SPEED