6-4
920512-768
Operator Manual
Routine Maintenance
9.0
FILTERS (Cont’d)
9.3
Replace drying pre-filter.*
1 X/year
9.4
Replace drying HEPA filter.*
1 X/year
9.5
Replace VHP blower in-line filter*
2 X/year
9.6
Verify VHP catalytic converter - Clean or replace as necessary.*
1 X/year
10.0
FINAL TEST
10.1
Clean dirt and lint from components. Verify all wiring, terminals and socket connec-
tions for damage or fraying.
Every
inspection
10.2
Remove all test equipment installed for inspection.
Every
inspection
10.3
Unlock electrical disconnect switch, position to ON and turn all utilities back ON.
Every
inspection
10.4
Run unit through two cycles to verify proper operation. Verify all displays and print-
outs.
Every
inspection
10.5
Install any panel or cover removed during inspection.
Every
inspection
10.6
Inspect work area and washer to ensure removal of all materials used during
inspection.
Every
Inspection
* Contact STERIS for this service. Service charges may be incurred. Consult your warranty for details.
Table 6-1. Preventive Maintenance Guide (Cont’d)
Recommended frequency of inspection is indicated in the right column. Usage and
utility conditions may require more or less frequent inspections.
Suggested
Frequency