10-44
Deposit Options for BSWSV, DPSV, LSSV and OSWSV
The Deposit Options dialog box in the Method Menu is used to control the
Deposition Potential. The default for this parameter is the initial potential for the
potential scan (Initial E). However, if some other potential is required for the
Deposition Potential, then the Deposit E option should be selected. The desired
Deposit E can then be entered. The electrode potential is always held at Initial E
during the Quiet Time.
Mode = BSWSV
General Parameters
Initial E (mV) = 0
-3276 to 3276
Final E (mV) = 0
-3276 to 3276
Rotation Rate (rpm) = 400
0 to 10000
Sensitivity = 1
µ
A/V
100 mA/V to 100 nA/V (10 pA/V with the
Low Current Module)
Deposit Time (sec) = 15
0 to 65535
Specific Parameters
Scan Rate (mV/s) = 4
1 to 200
S.W. Amplitude (mV) = 25
1 to 250
S.W. Frequency (Hz) = 14
5 to 250
Quiet time (sec) = 10
0 to 65535
Comments
a) The number of current samples averaged during each Step Period is
determined by the S.W. Frequency.
b) The Step Period is set at 1500 msec.
c) Sampling width is fixed at 1/4 cycle time period for all frequencies.
d) The sample interval is the product of the Scan Rate and the Step Period (e.g.,
the default value is 6 mV).
Mode = DPSV
General Parameters
Initial E (mV) = 0
-3276 to 3276
Final E (mV) = 0
-3276 to 3276
Rotation Rate (rpm) = 400
0 to 10000
Sensitivity = 1
µ
A/V
100 mA/V to 100 nA/V (10 pA/V with the
Low Current Module)
Deposit Time (sec) = 15
1 to 65535