Programming
69
Operating Instructions HYGROPHIL® H 4230-10 Serie A BA 030520
5.2.4.6
Reference pressure
Under exceptional operating conditions, the air pressure in the measuring
chamber may deviate from the pressure at the point where the measurement
gas is extracted. This is usually the case, for instance, if long gas sampling
hoses need to be laid. These differences in pressure give rise to minor
errors in the calculation of humidity values. For application cases where
such errors are undesirable, you can enter the pressure at the place of
measurement as a fixed value.
ENTER
Don’t make any changes
1
Enter and use a fixed reference pressure
0
Do not use a fixed reference pressure, go to the next
menu (sensor comparison)
ENTER
Don’t make any changes
Change reference pressure: enter new value in hPa using keyboard,
ENTER
(500...1500)
5.2.4.7
Sensor calibration
After installing or exchanging the temperature sensors, you must enter their
calibration data. You will find the calibration data on a plate at the sensor.
ENTER
Don’t make any changes
Change resistance value for TT at 0 °C: enter new value in hundredth
using keyboard,
ENTER
Example
TT at 0 °C = 100
entry: 10000
ENTER
Don’t make any changes
Change resistance value for TT at 80°C: enter new value in hundredth
using keyboard,
ENTER
ENTER
Don’t make any changes
Change resistance value for HT at 0 °C: enter new value in hundredth
using keyboard,
ENTER
ENTER
Don’t make any changes
Change resistance value for HT at 80 °C: enter new value in hundredth
using keyboard,
ENTER
EQUIPM.-CONFIG
SPN = 1013
EQUIPM.-CONFIG
TT 0°C = 10000
EQUIPM.-CONFIG
TT 80°C = 13089
EQUIPM.-CONFIG
HT 0°C = 10000
EQUIPM.-CONFIG
HT 80°C = 13089
EQUIPM.-CONFIG
SP-CORREC. ? 1/0