46
Troubleshooting
Problem Possible
Causes
Solutions
No electrical power to NANOpure
Ensure NANOpure DIamond power
DIamond.
cord is connected to a live power
source and completely plugged into
an electrical outlet as well as power
NANOpure DIamond
entry module on the unit. Make sure
completely inactive (pump and
on/off switch is in the “i” (on)
display not operating).
position.
Main fuse(s) blown.
Replace the main fuse(s) as
indicated in “Fuse Replacement.”
Power supply fuse blown.
Contact Customer Service
concerning power supply fuse
replacement.
Red caps in cartridge pack.
Remove red caps from cartridge
pack.
0.2 micron filter clogged.
Replace final filter as indicated in
Reduced or no product flow
“0.2 Micron Filter Replacement.”
from the dispense arm or
remote dispenser (if attached). Feed water not connected.
Reconnect feed water.
Failed solenoid.
Replace BK1190X2.
Pump protector (in reservoir), feed
Connect the pump protector, jumper
water line pressure switch or jumper
plug or pressure switch cord to the
plug not connected to pump interlock.
receptacle located on the upper left
hand section of the DIamond.
Display reads, “Check Inlet”
and pump does not run.
Insufficient feed water pressure or
Open feed water line shut off, con-
or volume to satisfy the NANOpure
nect feed water line to DIamond and
DIamond.
ensure sufficient pressure or suffi-
cient water level in the feed water
reservoir.
If you just started your DIamond,
Air in cell.
wait 1 minute to see if it corrects
Display reads, “...M
Ω
-cm”
itself.
Bad cell connection.
Check connection of the cell as it
attaches to the circuit board.
Recirculated water will not
Exhausted cartridge pack.
Replace cartridge pack as indicated
rinse up to desired purity level.
in “Cartridge Pack Replacement.”