Problem Possible
Causes
Solutions
PC board communication
Turn the unit OFF (I/O Switch) and restart.
Display reads “Meter Board
error.
Problem.”
System electronics failure.
Replace the resistivity meter board mounted
on the inside of the right door.
Wing head screws are not
Tighten wing head screws on unit manifold.
tight enough.
New cartridge pack: o-ring
Disconnect inlet water and depressurize
Water is leaking between the
missing from post (each of the system. Next, disconnect the manifold and
manifold and the cartridge
4 posts must contain an
verify/replace missing o-ring. Reinstall
pack.
o-ring).
cartridge pack as indicated in “Cartridge
Pack Replacement.”
Debris on o-ring or bottom of
Inspect o-rings and bottom of unit manifold.
unit manifold.
Remove any objects with a lint-free cloth.
Upon system startup, the
PC board communication
Turn the DIamond OFF (I/O Switch) and
message “TOC Comm. Error:
restart the unit.
(0A)” is displayed.
No TOC reading displayed.
TOC analyzer not activated.
Set “Use TOC Analyzer” to YES.
The display periodically reads
TOC lamp is burned out or
Replace TOC lamp.
“Replace TOC Bulb,” but TOC
intensity is insufficient.
(See “TOC Lamp Replacement”)
timer has not reached zero.
Display reads,
“<TOC Error>”
001
TOC memory error.
Cycle power, replace TOC Analyzer if it continues.
002
TOC communications error.
Cycle power.
005, 007
TOC sensor error.
Cycle power, replace TOC Analyzer if it continues.
006
Shorted TOC resistivity sensor.
Cycle power and place TOC analyzer in
hour long clean mode (under Diagnostics); if
still shorted, replace TOC Analyzer.
All other error codes.
Contact technical customer service.