Atmospheric Drain Connection
When the NANOpure DIamond TOC - UV flushes, the
water used is sent to drain through this connection. To
install:
1.
Locate the drain water tubing. This is the 1/4”
(.64 cm) O.D. tubing that is approximately 6 ft.
(1.9 m.) long. The atmospheric drain fitting is
located on the lower left side of the NANOpure
DIamond TOC - UV.
2.
Route the other end of the drain water tubing to
an atmospherically vented drain and make a
connection. Ensure there are no kinks in the
tubing and that it proceeds in a downward
plane. Proceed to “Initial Operation.”
13
I
NSTALLATION
Water Connections - Left side view
Caution
Do not connect feed water until direct-
ed to do so during “Initial Sanitization.”
Pump
Interlock
Remote 1
Remote 2
Feed
Water Inlet
Atmospheric
Drain