Wall Mounting
Install the NANOpure DIamond TOC - UV on a wall in a
convenient location that is accessible to water, an atmos-
pherically vented drain and electricity.
1.
Locate the wall bracket packed separately from
the unit.
2.
Using the wall bracket as a template, locate and
drill the mounting holes in the wall. A minimum of
four (customer-supplied) fasteners will be
required — two on the top and two on the bot-
tom.
3.
Attach the wall bracket to the wall using the cus-
tomer-supplied fasteners.
4.
Remove the locking screws on each side of the
wall bracket.
5.
Pull the two locking slides on each side of the
wall bracket out as far as they will go.
6.
Hang the unit on the wall bracket by sliding the
mounting pins into the wall bracket slots.
7.
Push the locking slides on each side of the wall
bracket in as far as they will go.
8.
Replace the locking screws.
Installing the Control Panel in a
Remote Location
For your convenience, the control panel can be removed
from the unit and mounted at a convenient location within
10 ft. (3.1 m.) of the unit. To remove the control panel
from the unit:
1.
Turn the unit OFF and disconnect it from the
power supply.
2.
Slide the control panel upward about 1/4” (.64
cm) to 3/8” (.95 cm) - (just enough to clear con-
nector on top cover), pull forward, and remove.
(The control panel fits tightly in the unit so it may
be difficult to remove.)
3.
Replace the control panel with the blank panel
provided with the unit to protect the electrical
connector.
11
I
NSTALLATION
Note
Please refer to the “General
Specifications” section for clearance
requirements.
Warning
Do not locate the NANOpure DIamond
TOC - UV directly over equipment that
requires electrical service. Routine
maintenance of this unit may involve
water spillage and subsequent electri-
cal shock hazard if improperly located.
NOTE
The outlet of a gravity feed storage
reservoir must be above or at the
same level as the inlet of the
NANOpure DIamond TOC - UV.
********************
************************
Caution
Wall composition, condition and con-
struction, as well as fastener type,
must be considered when mounting
this unit. The mounting surface and
fasteners selected must be capable of
supporting a minimum of 150 lbs.(68
kg). Inadequate support and/or fasten-
ers may result in damage to mounting
surface and/or equipment. If you are
unsure of mounting surface composi-
tion, condition and construction or cor-
rect fasteners, consult your building
maintenance group or contractor.
Содержание NANOpure DIamond D11951
Страница 52: ...Wiring Diagram 52...
Страница 55: ...55 TITLE RIGHT...