9
Choosing a Site
The NANOpure DIamond TOC - UV features a removable
control panel display which allows the system to be
mounted almost anywhere within 10 ft. (3.1m) of the unit.
Use the wall bracket for wall mounted systems as a tem-
plate to drill mounting holes. (The NANOpure DIamond
TOC - UV does not include screws and fasteners for
mounting.) Please refer to the “General Specifications”
section for clearance requirements.
Installation
NOTE
The outlet of a gravity feed storage
reservoir must be above or at the
same level as the inlet of the
NANOpure DIamond TOC - UV.
********************
************************
Caution
Wall composition, condition and
construction as well as fastener type
must be considered when mounting
this unit. The mounting surface and
fasteners selected must be capable of
supporting a minimum of 150 lbs.(68
kg). Inadequate support and/or fasten-
ers may result in damage to mounting
surface and/or equipment. If you are
unsure of mounting surface
composition, condition and construc-
tion or correct fasteners, consult your
building maintenance group or
contractor.
Warning
Do not locate the NANOpure DIamond
TOC - UV directly over equipment that
requires electrical service. Routine
maintenance of this unit may involve
water spillage and subsequent electri-
cal shock hazard if improperly located.
Do not use in the presence of
flammable or combustible materials;
fire or explosion may result. This
device contains components which
may ignite such materials.
Содержание NANOpure DIamond D11951
Страница 52: ...Wiring Diagram 52...
Страница 55: ...55 TITLE RIGHT...