27
3.
Press the UP or DOWN arrow to select “Yes.”
4.
Press ENTER and the timer will be reset for
1820 hours of operational time.
Press BACK two times to return to the (Idle) mode.
Resetting the TOC Lamp Timer
When the NANOpure DIamond TOC - UV is in an (Idle)
position:
1.
From the (Idle) mode, when display reads,
“NANOpure DIamond (Idle) xx:xx:xx
xx/xx/xx,” press the UP or DOWN arrow until the
display reads, “Reset Timers?” Press ENTER.
2.
Press ENTER until the display reads, “Reset
TOC Timer.” It will also read, on the right side of
the second line, the hours of usage which have
occurred since the TOC Timer was last reset.
3.
Press the UP or DOWN arrow to select “Yes.”
4.
Press ENTER and the timer will be reset for
1300 hours of operational time.
The display will now display “Reset San Timer.” Press
BACK two times to return to the (Idle) mode.
Using the TOC Analyzer
(Factory set to “No.”)
This feature is accessible from the (Idle) mode.
Use the analyzer when TOC values and resistivity/con-
ductivity readings are to be displayed. DO NOT use the
analyzer when only resistivity/conductivity readings are to
be displayed.
1.
From the (Idle) display press UP or DOWN
arrow until “Advanced Menu?” is displayed.
2.
Press ENTER.
3.
Press ENTER when “Printer Attached?” is dis-
played.
4.
Continue pressing ENTER until “Use TOC
Analyzer?” is displayed.
5.
Press UP or DOWN arrow to choose “Yes.”
U
SER
S
ETTINGS