23
Open the carton carefully making certain that all parts are accounted for
before packaging materials are discarded. After unpacking, if damage is found,
promptly report it to the carrier and request a damage inspection promptly.
IMPORTANT: Failure to request an inspection of damage within a few
days after receipt of shipment absolves the carrier from any liability for damage.
You must call for a damage inspection promptly.
LOCATION:
Place the unit where it will be used, away from external vibration sources,
drafts and wide variations in ambient temperature. Choose a location near a
power supply that matches the unit nameplate requirements.
CAUTION:
THIS UNIT IS DESIGNED FOR USE WITH GASEOUS AIR, CO
2
OR
NITROGEN ONLY. DO NOT USE WITH ANY OTHER GASES. FAILURE TO
OBSERVE THESE PRECAUTIONS CAN RESULT IN EXPLOSION AND/OR
FIRE AND SERIOUS INJURY AND/OR DEATH AND PROPERTY DAMAGE.
LEVELING:
Level the unit by using a bubble-type level placed on a shelf. Turn the
leveling feet clockwise to raise and counterclockwise to lower the height of the
respective legs of the Incubator. The unit should be level from front to back and
side to side.
ELECTRICAL REQUIREMENTS:
120 VAC models require a 120 VAC, 50/60 Hz power source. They are
supplied with a 3-wire line cord that should be plugged into an outlet designed for
3-prong plugs. If an extension cord is used, it also should be the 3-wire grounded
type. For an outlet designed to accept 2-prong plugs, ungrounded, it is
required that a qualified electrician replace the outlet with a new grounded type.
240 VAC models require a 240 VAC, 50/60 Hz power source. Because of
the variety of plug configurations in use worldwide for 240 VAC power, the unit is
furnished with the plug removed. The user must install a plug to conform with
local code and configuration requirements.
100 VAC models require a 100 VAC, 50/60 Hz power source. Because of
the variety of plug configurations in use worldwide for 100 VAC power, the unit is
furnished with the plug removed. The user must install a plug to conform with
local code and configuration requirements.
If a plug must be installed, use only the 3-prong grounded type, rated for
the unit load requirements and matching the power outlet. Make sure the green
ground wire is secured to the plug ground terminal.
NOTE:
LEAVE UNIT DISCONNECTED WHEN NOT IN USE.
INSTALLATION:
(Con’t)
SHELVING/PLENUM INSTALLATION-SELF-SUPPORTING SYSTEM COMPONENTS,
TOP PLENUM
SHELF SUPPORT SIDE PANELS