CONTROL PANEL: (Con’t)
SET
CAL
CO2
TEMP
DOOR
SCAN
ºC
%CO2
HEAT
HOLD
HI LIMIT SAMPLE
1
2
3
4
5
6
7
8
9
10 11
12
4.
CO
2
COMMAND KEY: Pressing the
SET
key and then the
CO
2
key
allows the user with the Up and Down keys to enter the desired carbon
dioxide level in the Incubator.
•
High and low limit alarm set points are factory set at ±1% above and
below the user-established set point. In the event CO
2
exceeds the
indicated limits, an audible alarm will sound and indicating light
above the key will flash.
•
Pressing any key will silence the alarm for a period of one hour. The
user should immediately investigate the reason for the alarm and
take steps to correct the problem. Note that the indicating light
continues to flash. If a problem is not solved the alarm resumes.
•
By pressing the CAL (#3) and CO
2
(#4) keys sequentially users can
calibrate CO
2
. The Up and Down arrow keys will change the readout
to correspond with the readings obtained from known standards of
CO
2
measurement.