17
CONTROL PANEL: (Con’t)
SET
CAL
CO2
TEMP
DOOR
SCAN
ºC
%CO2
HEAT
HOLD
HI LIMIT SAMPLE
1
2
3
4
5
6
7
8
9
10 11
12
10.
HI-LIMIT THERMOSTAT: This is a backup temperature controller
that takes charge in the event the microprocessor fails. The hi-limit
thermostat will then control the temperature at slightly above the set point.
11.
HI-LIMIT STATUS LAMP: Indicates when the hi-limit is active.
12.
SAMPLE PORT: This access port is available for sampling the
atmosphere in the chamber when executing calibration procedure such as
might be done using a Fyrite or similar chemical-based CO
2
analysis
procedure.
13
13.
LED RH DISPLAY; MODELS 460RH, 465RH, 490RH, 495RH
ONLY
: With
the toggle switch in the up position, actual chamber humidity is displayed as
% RH. Placing the toggle switch located on the incubator’s back panel, in
the down position, the RH alarm set point is displayed, and the audible
portion of the alarm is disabled.
FEATURES:
(Con’t)
WATER-JACKETED MODELS:
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com