21
DOOR AJAR WARNING: In case the door is inadvertently left open beyond a 5
minute period, an audible beep is sounded and is synchronized with the flashing
of the DOOR-HEAT indicator light to warn the user to correct the condition.
AUDIBLE AND VISUAL ALARMS: The following table summarizes the audible
and visual alarms that are initiated upon the occurrence of various events.
AUDIBLE
FLASHING
FLASHING
BEEP
LIGHT
LED
ABOVE KEY
DISPLAY
Temp (exceeds limits):
YES
YES
NO
CO
2
(exceeds limits):
YES
YES
NO
Door Ajar:
YES
YES
NO
Power Out:
NO
NO
YES
Low Water, 460/465 SERIES:
NO
STEADY
NO
Press any key to silence alarm that will remain silent for a period of one
hour. However, the flashing light will continue. If the condition initiating the alarm
remains uncorrected at the end of that time, the audible alarm will resume.
PASSWORD:
In the event that you want to prevent unauthorized access to the operation
of the unit, a password feature is available.
FEATURES:
(Con’t)
OPTIONAL ALARM RELAY OUTPUT: REAR OF UNIT
The incubator incorporates an output on the rear panel for use as a central alarm
voltage-free (“dry contacts”) connection port. This output may be tied in with any central
alarm system to give notice of any alarm condition: High or Low Temperature, High or
Low CO2, Low Water Level (water jacket models), Power Outage or Door Opened.
A standard relay is used and is normally energized while the incubator is in
operation. Common, Normally Open, and Normally Closed pin outputs are routed to the
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com