42
OPERATION:
(Con’t)
RS232 STATISTICS FOR 465/495 SERIES INCUBATORS: (Con’t)
•
The output format used is as follows:
PRINTER (CTS USE): DOUBLE LINE:
⇒
Temp =
XXX.X C(LF)(CR)
<<
Line Feed/Carriage Return
⇒
CO2 =
XX.X% (LF)(CR)(LF)(CR)
•
Capture Program (PC): Single Line:
⇒
Temp =
XXX.X C, CO2 = XX.X% (LF)(CR)
•
Upon resumption of power after a power outage or any power-up event:
Single Line:
⇒
...RESUME SYSTEM POWER (LF)(CR)
•
The RS232 data pins available at the rear of the unit (D-subminiature 9-pin)
are:
⇒
Signal Ground: Pin 5 (To your device ground)
⇒
Transmit (TX): Pin 3 (To your device "Receive data" input)
⇒
CTS (RX): Pin 8 (From your device CTS, RTS or DTR output)
•
Characters are transmitted in an 8-bit UART fashion:
⇒
>>
10 bits are transmitted: a start bit (0), 8 data bits (LSB) and a stop bit.
⇒
NO additional parity bits are included and the MSB data bit is always
zero.
⇒
>>
Meets RS232C and V.24 Electrical Specifications.
POWER MAINS PROTECTION: MODELS 460-1CE, 465-1CE, 490-1CE & 495-1CE
If a fuse blows, have a qualified person replace it with a properly rated
fuses—these now follow here: The internal primary of the transformer fuse is:
0.6 AMP, Littlefuse
#313.006, Slo-Blo Type (Lab-Line Part #330-370-00);
internal motor lead fuse is: 1 AMP, Littlefuse
#313.001 (Lab-Line Part #330-
369-00).
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com