SECTION 6
OPERATION
DANGER:
DO NOT USE IN THE PRESENCE OF FLAMMABLE OR
COMBUSTIBLE MATERIALS OR EXPLOSIVE GASES. DO NOT USE IN THE
PRESENCE OF PRESSURIZED OR SEALED CONTAINERS—FIRE OR
EXPLOSION MAY RESULT, CAUSING DEATH OR SEVERE INJURY.
WARNING:
DO NOT HEAT ANY SUBSTANCE ABOVE A TEMPERATURE
WHICH WILL CAUSE IT TO EMIT TOXIC FUMES—DEATH OR SEVERE
INJURY MAY RESULT.
DANGER:
USE ONLY AN INERT GAS SUCH AS CARBON DIOXIDE IN THE
INCUBATOR. DO NOT
UNDER ANY CIRCUMSTANCE
INJECT OXYGEN OR
OTHER EXPLOSIVE GAS OR MIXTURE INTO UNIT. FAILURE TO OBSERVE
THESE PRECAUTIONS CAN RESULT IN EXPLOSION AND/OR FIRE AND
SERIOUS INJURY OR DEATH TO PERSONNEL AND PROPERTY DAMAGE
START-UP:
•
Connect unit to power supply meeting the requirements as noted on the
nameplate.
•
Turn on the Incubator with rocker switch that is located on back left side.
•
All display segments and parameter lamps should light for 5 seconds.
OPERATING BASICS:
•
Press
the
SET
key first and then the parameter—
CO
2
, TEMP, DOOR
HEAT
—which is to be adjusted.
•
Use the UP and DOWN arrow keys to enter the desired value.
•
Wait 5 seconds and the value is entered automatically into the unit's
nonvolatile memory. Storage is confirmed via 2 short beeps at the end of 5
seconds.
•
If you access a parameter and initiate no action within the time frame of 10
seconds, the unit times out and returns to the prior state.
•
Each press of the
SCAN/HOLD
key is programmed to follow a sequential
pattern as follows: HOLD temperature (for continuous viewing), HOLD CO
2
(for continuous viewing) and alternate scanning of temperature and CO
2
.
•
To view your previously established set point for any parameter, press
SET
and the
TEMP, CO2
or
DOOR HEAT
key(s)—unit will revert to its normal
display after 10 seconds.