32
INSTALLATION:
(Con’t)
INITIAL FILLING OF WATER JACKET, 460 SERIES:
1. Turn unit on. The low-water LED will display a low-water message.
2. To fill the water jacket, open the unit’s front door and locate water inlet, top left
under front control panel.
FRONT/LEFT OF UNIT:
LED READOUT DISPLAY
WATER INLET (FEMALE)
QUICK-CONNECT RESET BUTTON
QUICK-CONNECT
WATER FITTING (MALE)
SIDE VIEW
FLEXIBLE HOSE
3. Verify overflow outlet on rear of unit is not obstructed. Connect a ¼” (6.35 mm) ID
piece of tubing to overflow outlet and place other end of tubing in a 1 liter container
to catch any possible overflow.
4. Attach appropriately sized tubing from distilled water source to the water inlet.
5. Turn distilled water source on. Flow rate should not be greater than 1 liter/minute.
6. Monitor low-water light on control panel until it goes out and
immediately stop filling procedure—turn off the distilled water source WITHIN THE
NEXT FEW SECONDS, all excess water will run from the overflow outlet located
at the rear of the unit.