Site~Rite
®
8 Ultrasound System
Page 78
Quality of Service (QoS) – Bluetooth
The Bluetooth™ functionality for the Site~Rite
®
8 Ultrasound System operates on a best efforts basis, since latency in
data transmission will not affect critical ultrasound imaging. As these connections are point to point, they do not require or
rely on a formal QoS WLAN. The link capacity between devices is designed to provide the necessary bandwidth for
proper data transmission, without competition for resources between devices.
To ensure a proper connection, keep the separation distance between Bluetooth™ devices and the Site~Rite
®
8
Ultrasound System console as short as possible. The maximum operating distance for Bluetooth™ technology is < 10
meters. Bluetooth™ devices generally perform better at shorter distances however, on the order of ~3 meters, with a
direct line of sight preferable.
IEEE 802.11b/g/n Wireless Security
To guard against unauthorized access to network resources, ensure that the Site~Rite
®
8 Ultrasound System is
connected to a secure network. Use of WPA2 security is preferred over older protocols, such as WEP, as it provides a
more secure network.
Contact your local IT administrator for more information on which networks can be used and what
settings should be used.
Bluetooth™ Security
The Site~Rite
®
8 Ultrasound System employs Bluetooth™ connectivity for use with approved accessories. Bluetooth™
technology does not provide rigorous security measures to prevent unauthorized access, as wifi does. As such,
appropriate measures should be taken to ensure that access is limited to approved accessories to be used with the
system.
Wireless coexistence issues and mitigations:
There is a potential for wireless conflicts, as 802.11b/g/n and Bluetooth™ both operate on the 2.4 GHz frequency band,
as well as interference from other RF devices. If you encounter problems connecting to a network or Bluetooth™ device,
remove or separate possible interfering sources as much as possible. Objects and people obstructing line of sight
connections between wireless devices can also cause interference. Maintain a clear path between the Site~Rite
®
8
Ultrasound System and paired accessories as much as possible to ensure highest performance.
An issued or revision date for these instructions is included for the user’s information. In the event that two years have lapsed
between the issued revision date and use of the product, contact Bard Access Systems, Inc. to see if any additional product
information is available.
Revision date: February 2020
Manufacturer:
Bard Access Systems, Inc.
605 North 5600 West
Salt Lake City, UT 84116
U.S.A.
Customer Service: 1-800-545-0890
Technical/Clinical Support: 1-800-443-3385
Fax: 1-801-522-4948
www.bardaccess.com
Bard, Cue, Pinpoint, Sherlock, Sherlock 3CG, and Site~Rite are trademarks and/or registered trademarks of C. R. Bard, Inc. All
other trademarks are the property of their respective owners.
© 2020 C. R. Bard, Inc. All rights reserved.
0752435 2001R
Содержание Site-Rite 8
Страница 1: ...Instructions for Use...