9. CUE™ NEEDLE TRACKING SYSTEM MODE (if enabled)
9.1. Enabling Cue™ Needle Tracking System
Ensure the Cue™ Probe is attached (see Section 3.2). Turn on the system and ensure the ultrasound system is scanning.
The Cue™ Activator icon will appear, as shown below, when the Cue™ Probe is attached. Turn on the system and
ensure the system images. Connect the provided USB cable to the Cue™ Activator. Then connect the other end of the
USB cable to a USB port on the ultrasound system.
Ensure the Activator icon indicates a connection.
9.2. Activating the Cue™ Needle Tracking System Compatible Needle
Place a tray containing a Cue™ Needle Tracking System
compatible needle on the Cue™ Activator as shown.
Note:
Ensure the tray is fully seated in the Activator,
making complete contact with the Activator.
When the needle is activated, the following window shall appear.
The Cue™ Needle Tracking System compatible needle and software are now activated.
Not Connected
Connected
Needle Information
If selected, this window will
not be shown for the
currently selected Clinician
(see Section 13).