Site~Rite
®
8 Ultrasound System
Page 21
3.
Place the Catheter in the middle of the
vessel by pressing and holding the catheter
icon and dragging it to the desired location.
4.
Evaluate whether or not the vessel complies with the user specified rule.
If the Vessel Occupancy Rule Circle
is
smaller
than the vessel, the selected catheter
complies with the user-specified rule.
If the Vessel Occupancy Rule Circle is
larger
than the
vessel, the selected catheter does
NOT
comply with the
specified rule. Select a smaller catheter, or find a larger
vessel to comply with the user-specified rule.
Vessel Measurement Indicator
This icon indicates whether the vessel
occupancy is calculated by
Diameter
or
Area
. This can be set in the
Ultrasound Settings Tab
(see Section 11.1).
Diameter
Area
Catheter Icons
Toggles through available
catheter icon sizes
Select desired catheter
icon size.
Vessel Occupancy Percentage
This is the user-defined vessel
occupancy percentage rule. The
vessel occupancy percentage is the
percent of the vessel occupied by the
catheter as defined by the clinician.
This rule can be set in the Ultrasound
Settings Tab (see Section 11.1).
Vessel Occupancy
Rule Circle
The dotted circle represents
the rule specified in the
ultrasound settings
(see Section 11.1).
Catheter Icon
Represents the size of the
selected catheter.
Содержание Site-Rite 8