Site~Rite
®
8 Ultrasound System
Page 76
System Accuracy with Needle Guides
Measurement
Error
Range‡
Vertical
≤ 1.5mm
5–35mm
Horizontal
≤ 1.5mm
5–35mm
‡ Available depth range of the Site~Rite
®
Needle Guide and Pinpoint™ GT Needle Guide
Measurement Tool Error
Measurement
Error
Range
Area
< 10%
0–804mm
2
Diameter
< 10%
0–32mm
Area Ratio
< 10%
0–100%
Diameter Ratio
< 10%
0–100%
Cue™ Needle Tracking System Accuracy
Accuracy is within the Intersection Window*
* Results have 95% confidence with 99.4% reliability from bench top testing in a controlled environment.
Caution:
Improper technique and environmental conditions may introduce variation in accuracy.
Pinpoint™ GT Needle Technology System Accuracy
±1.45mm*
* Results have 95% confidence with 96% reliability from bench top testing in a controlled environment.
Caution:
Improper technique and environmental conditions may introduce variation in accuracy.