Page 62
Troubleshooting – Site~Rite
®
8 Ultrasound System
SYMPTOM
DESCRIPTION
SOLUTIONS
Error Code
EC:111
EC:113
EC:199
System Error
1.
Restart the system.
Note:
If error persists, call Customer Service at 1-800-545-0890. An alternate
number may be available in the System Information Tab (Section 11.5).
Error Code
EC:112
File Reading Error
2.
Failed to read configuration file. Some user settings may be lost.
Note:
If error persists, call Customer Service at: 1-800-545-0890. An alternate
number may be available in the System Information Tab (see Section 11.5).
Error in WiFi Networks List
WiFi Connection Error
1.
Ensure the security key is correct.
2.
Ensure WiFi configuration is correct.
3.
Contact your network administrator.
Note:
If error persists, call Customer Service at: 1-800-545-0890. An alternate
number may be available in the System Information Tab (see Section
11.5).
Screen not oriented
properly or cut off when
HDMI cable is plugged in
1.
Plug the USB keyboard.
2.
With the system in the main ultrasound scree, press “ctrl+shift+right
arrow”. The screen should rotate.
3.
Press “ctrl+shift+q” to restart the application.
This must be repeated every power cycle.
Patient/Clinician Backup Error
1.
System is approaching file capacity limit. Backup cannot be performed.
Consider removing patient data or backups in order to free up space.
Note:
This device is equipped with a lithium ion battery. Lithium ion battery performance is known to degrade over time based on environmental and
use conditions. These messages are intended to help manage the life of the battery.
Pre-Critical Warning
This window appears when the battery is approaching the end of its useful life.
It is recommended that the battery be replaced.
Critical Warning
This window appears when the battery’s integrity has reached a critical stage,
indicating the system detected the battery has reached the end of its useful life
and that without AC power, the system is likely to shut down. The message will
only disappear at startup when the system is connected to AC power.