16. TROUBLESHOOTING
16.1.
Site~Rite® 8 Ultrasound System Troubleshooting
Troubleshooting – Site~Rite
®
8 Ultrasound System
SYMPTOM
DESCRIPTION
SOLUTIONS
Error Code
EC:201-203
EC:205-206
EC:211
Ultrasound Hardware Failure
Reboot the system.
Note:
If error persists, call Customer Service at: 1-800-545-0890. An alternate
number may be available in the System Information Tab (Section 11.5).
Error Code
EC:204
Probe Disconnect Error
If applicable, ensure the probe is securely connected to the system. See
Section 3.3.
Note:
If error persists, call Customer Service at: 1-800-545-0890. An alternate
number may be available in the System Information Tab (Section 11.5).
Error Code
EC:101
System temperature too high
Shut down the system and allow to cool.
Note:
If error persists, call Customer Service at: 1-800-545-0890. An alternate
number may be available in the System Information Tab (Section 11.5).
System plays a sound
every minute.
Low Battery
Connect system to power for continued operation and to recharge battery.
Error Code
EC:110
Battery Error
Connect AC adapter for operation. Ensure the battery switch is in the “
”
position
(Section 4).
Note:
If error persists, call Customer Service at: 1-800-545-0890. An alternate
number may be available in the System Information Tab (Section 11.5).
Power Problem
System will not turn on or turns
on but immediately turns off
1.
Press and hold the power switch for at least 2 seconds to turn the
system on.
2.
Connect the system to AC power for operation and to recharge the
battery.
Note:
If error persists, call Customer Service at: 1-800-545-0890. An alternate
number may be available in the System Information Tab (Section 11.5).
A/C adapter indicator
does not change states
when A/C power is
connected
Connection Error
1.
Unplug the AC adapter and plug it back into the system.
2.
Restart the system if necessary.
Note:
If error persists, call Customer Service at: 1-800-545-0890. An alternate
number may be available in the System Information Tab (Section 11.5).
System Does Not Charge
Battery Charging Failure
1.
Unplug the AC adapter and plug it back into the system.
2.
To charge the battery in temperatures exceeding 32°C, the system may
need to be turned off.
Note:
If error persists, call Customer Service at: 1-800-545-0890. An alternate
number may be available in the System Information Tab (see Section 11.5).
Error Code
EC:102
USB Storage Device Error
The USB storage device may be full or may not be a compatible USB storage
device. Replace the USB storage device.
Note:
If error persists, call Customer Service at: 1-800-545-0890. An alternate
number may be available in the System Information Tab (see Section 11.5).
Error Code
EC:103
Hard Drive Error
Restart the system.
Note:
If error persists, call Customer Service at: 1-800-545-0890. An alternate
number may be available in the System Information Tab (see Section 11.5).