Site~Rite
®
8 Ultrasound System
Page 32
10.3.
Pinpoint™ GT Needle Technology Needle Tracking
Once Pinpoint™ GT Needle Technology has finished calibrating, the system is ready to begin tracking Pinpoint™ GT
Needle Technology compatible needles.
Caution:
Movement of the probe during calibration may result in calibration error.
Caution:
Keep the needle and any ferromagnetic objects, e.g., wired undergarments, metal instruments, watches,
jewelry, electronics metal bed rails, etc. at least 2 feet away from probe when calibrating. Failure to do so
may impact the accuracy of needle tracking.
Caution:
Equipment operating in close proximity may emit strong electromagnetic or radio frequency interference
which could affect the performance of this device. Avoid operating the device near pumps, cauterizers,
diathermy equipment, cellular phones, or other portable and mobile radio frequency communications
equipment. Maintain equipment separation of at least 5 feet (1.5m).
Caution:
Excessive movement of the probe after calibration may result in reduced needle tracking accuracy.
Calibration Status Icon
Once Pinpoint™ GT
Needle Technology is
calibrated, the status icon
will change to a green
check mark.
Freeze Button
After calibration has
completed, the freeze
button is accessible.
Calibration Status
Border
This border indicates the
calibration status of
Pinpoint™ GT Needle
Technology.
Green Border –
“READY… LIMIT
PROBE MOVEMENT”
System is calibrated and
ready for needle tracking.
Calibrate Button
After calibration has
completed, the calibrate
button is accessible.