Proceed with insertion – watch needle advancement onscreen. The solid green lines represent the needle
shaft and tip. The green square represents the area where the needle will intersect the ultrasound plane
e.
During advancement, maintain the Intersection Window centered with the intended target structure
f.
When the solid green lines reach the Intersection Window, it means the needle has intersected the plane of
the ultrasound beam within the Intersection window. It does not mean insertion has been successful or the
structure of insert has been reached.
g.
When attempting vessel access, watch for:
•
Anterior vessel wall indenting (once the puncture occurs the vessel will return to normal shape)
•
Blood return in the device
•
Needle reflection 9flash of the needle tip) on the ultrasound image
h.
Complete Procedure per suggested institutional guidelines
5b.
In-Plane
Ultrasound Guided Insertion Using the Cue™ Needle Tracking System
a.
Retrieve the activated needle and move it towards the side of the probe at a 45° angle until it is within 0.5cm
from the probe. Ensure the Cue™ Needle Tracking System virtual needle overlay appears on the display and
follows needle movement
Caution
:
If needle tracking is not functioning properly, discontinue use of Cue™ Needle Tracking System.
b.
Position the needle by adjusting the insertion angle and distance from the probe to align the Cue™ Needle
Tracking System Overlay with intended target structure
Note:
In-Plane Mode does not have an Intersection Window.