A151 Benchtop Conductivity Meter
7
Temperature Coefficient
The A151 conductivity meter contains two temperature correction methods. The linear temperature correction is appropriate for most samples. If the current
sample is belong to the natural water (e.g., natural ground, well, or surface waters), using the non-linear correction is necessary. Note: non-linear correction
can only be performed at temperature range from 0° C to 36° C. If the temperature value is out of above range, the meter will show a warning.
To calculate the temperature coefficient,
please refer to addendum “How to calculate the temperature coefficient”.
Pure Water Coefficient
The pure water coefficient is used to correct the samples with a conductivity of less than 5µ S/cm. If enabled, the meter will be automatically calculated and
applied coefficient for ultra-pure water measurements.