A151 Benchtop Conductivity Meter
13
Multi-point calibration
2.1
Ensure that you have selected 2 or 3 points calibration in the setup menu.
2.2
Repeat steps 1.2 to 1.6
above. When the first calibration point is completed, the display will show “Calibration Point 2”. The meter prompts you to
continue with second point calibration.
2.3
Rinse the conductivity electrode with distilled water. Place the electrode into the next standard solution, the meter automatically shows current
calibration standard (e.g., 12.88mS/cm).
2.4
If necessary, press the
or
key to modify the calibration value.
2.5
Press the
Enter
key, the meter begins the calibration.