A151 Benchtop Conductivity Meter
12
1.4
Place the conductivity electrode into the standard solution, the meter automatically shows current calibration standard (e.g., 1413µ S/cm).
1.5
If necessary, press the
or
key to modify the calibration value.
1.6
Press the
Enter
key, the Calibrating.. icon shows on the display.
1.7
Wait for the reading
to stabilize, the meter automatically shows “Calibration is completed” and returns to the measurement mode.