A151 Benchtop Conductivity Meter
19
Conductivity Electrode Care and Maintenance
•
DO NOT touch the measurement area of electrode and always make sure that is clean.
•
If there is a build-up of solids inside the measurement area, these should be removed very carefully with a cotton bud soaked in solvent, taking care
not to touch the metal parts of the inner cell. After the solids are removed, the electrode must be recalibrated.
Specifications
Conductivity
Model
A151
Range
0.01~20.00, 200.0, 2000µ S/cm, 20.00, 200.0mS/cm
Accuracy
± 0.5% F.S
Resolution
0.001, 0.01, 0.1, 1
Calibration Points
1 to 5 points
Calibration Solutions
10µ S/cm, 84µ S/cm,1413µ S/cm,12.88mS/cm,111.8mS/cm
TDS
Range
0.00mg/L~100.0g/L (Max. 200g/L)
Accuracy
± 1% F.S
TDS Factor
0.01~1.00 (Default 0.5)
Salinity
Range
0.00~80.00ppt, 0.00~42.00psu, 0.00~8.00%
Accuracy
± 1% F.S
Measurement Modes
Practical Salinity (psu), Natural Seawater (ppt) or %
Resistivity
Range
0.00~30.00M
Ω
Accuracy
± 1% F.S
Resolution
0.01, 0.1, 1
Temperature
Range
0~105° C, 32~221° F
Accuracy
± 0.5° C
Calibration Points
1 point
General
Temperature Compensation
0~100° C, 32~212° F, Manual or Automatic
Temperature Coefficient
Linear, Non-linear and Pure water compensation
Normalization Temperature
20° C or 25° C
Cell Constant
2-pole electrodes (K=0.1, 1, 10) or 4-pole electrode
Memory
Stores up to 1000 data sets
Output
USB Communication Interface
Power Requirements
DC12V/2A, using AC adapters, 220VAC/50Hz
Dimensions
240 (L) × 220 (W) × 80 (H)mm
Weight
1.7kg