A151 Benchtop Conductivity Meter
10
Prior to Use
Soak the conductivity electrode for a few minutes in tap water to remove dirt and oil stains on the electrode.
Temperature Compensation
For better accuracy, we recommend the use of either a sensor with a built-in or a separate temperature probe for the calibration or measurement.
Automatic Temperature Compensation
Connect the temperature probe to the meter (Refer to page 4 “Connecting the Temperature Probe”). The ATC icon immediately app
ears on the display, the
meter is now switched to the automatic temperature compensation mode.
Manual Temperature Compensation
If the meter does not detect a temperature probe, the MTC icon will show on the display indicating that the meter is switched to the manual temperature
compensation mode. To set the temperature value, follow the steps below.
1.
Press and hold the
o
C
key for 3 seconds to enter the temperature setting mode.
2.
Press the
or
key to modify the temperature value.
3.
Press the
Enter
key to confirm, the meter returns to the measurement mode. Setting is completed.
Press the
or
key once, the setting value will increase or decrease by 0.1. Press and hold the
or
key, the setting value will increase or
decrease by 1.