A151 Benchtop Conductivity Meter 

11 

 

Selecting the Conductivity Electrode 

The A151 conductivity meter is capable of using the 3 types of 2-pole conductivity electrodes. Prior to the calibration and measurement, you need to select 
the different sensors as per your sample concentration. The table below shows available electrodes and its effective measuring ranges.   

 

CONDUCTIVITY ELECTRODE 

MEASURING RANGES 

CELL CONSTANT 

CON-0.1 

0.1~100µ S/cm 

K=0.1 

CON-1 

10µ S/cm~10mS/cm 

K=1 

CON-10 

100µ S/cm~200mS/cm 

K=10 

 
If the 4-pole conductivity electrode is selected, its best measurement range is 100µ S/cm to 200mS/cm. 
 

Conductivity Calibration 

The meter allows 1 to 3 points calibration in the conductivity mode. To ensure higher accuracy, we recommend that you perform 3 points calibration or select 
a standard value close to the sample value you are measuring. The meter will automatically detect these conductivity standard solutions and prompt the 
user to calibrate the meter. When the calibration is completed, all new calibration values will automatically override existing data.   
 
The following table shows acceptable conductivity range of standard solution for each measuring range. 

 

MEASURING RANGE 

CALIBRATION SOLUTION RANGE 

DEFAULT 

0~20µ S/cm 

7~17µ S/cm 

10µ S/cm 

20~200µ S/cm 

70~170µ S/cm 

84µ S/cm 

200~2000µ S/cm 

700~1700µ S/cm 

1413µ S/cm 

2~20mS/cm 

7~17mS/cm 

12.88mS/cm 

20~200mS/cm 

70~170mS/cm 

111.8mS/cm 

 

Single point calibration 

1.1 

Ensure that the meter is in the conductivity measurement mode and you have selected 1 point calibration in the setup menu. 

1.2 

Rinse the conductivity electrode with distilled water, then rinse with a small amount of standard solution. 

1.3 

Press the 

Cal

 

key, the display shows “Calibration Point 

1

 and waits for recognizing the standard solution. 

 

 

 
 
 

Содержание A151

Страница 1: ...A151 Benchtop Conductivity Meter Instruction Manual BANTE INSTRUMENTS CO LTD...

Страница 2: ...ent meets following conditions Relative humidity is less than 80 Ambient temperature is greater than 0 C and less than 60 C No potential electromagnetic interference The following list describes the s...

Страница 3: ...seconds Locks the measured value Resume measuring Sends data to a printer or computer Stores current reading to memory Increase value or scroll up through the menu item Views the calibration report o...

Страница 4: ...e arm automatically rises or falls you need to adjust the screw until arm locate at any position 1 Remove the plastic cover from the electrode arm 2 Use the screwdriver to tighten the screw moderately...

Страница 5: ...and dry Connecting the Temperature Probe 1 Place the temperature probe into the circular hole of the electrode arm 2 Insert the phone plug to the connector socket labeled C Ensure the connector is fu...

Страница 6: ...temperature compensation type and coefficient Linear 2 1 C Non linear Pure Water Coefficient Enable When the option is enabled pure water coefficient will be applied automatically for ultra pure water...

Страница 7: ...the default parameter The meter provides two methods for parameter settings Press the or key to modify the value press the Enter key to confirm If the cursor appears below the first digit press the or...

Страница 8: ...n is necessary Note non linear correction can only be performed at temperature range from 0 C to 36 C If the temperature value is out of above range the meter will show a warning To calculate the temp...

Страница 9: ...re is used to lock a measurement endpoint If enabled the meter will automatically sense a stable reading and lock the measurements The HOLD icon appears on the display Press the Meas key the meter res...

Страница 10: ...y and wait for entering the correct digits Press the or key to input the password press the Enter key to confirm Once you have successfully entered the setup menu selecting the Disable The password wi...

Страница 11: ...the meter is now switched to the automatic temperature compensation mode Manual Temperature Compensation If the meter does not detect a temperature probe the MTC icon will show on the display indicat...

Страница 12: ...lect a standard value close to the sample value you are measuring The meter will automatically detect these conductivity standard solutions and prompt the user to calibrate the meter When the calibrat...

Страница 13: ...ically shows current calibration standard e g 1413 S cm 1 5 If necessary press the or key to modify the calibration value 1 6 Press the Enter key the Calibrating icon shows on the display 1 7 Wait for...

Страница 14: ...ed the display will show Calibration Point 2 The meter prompts you to continue with second point calibration 2 3 Rinse the conductivity electrode with distilled water Place the electrode into the next...

Страница 15: ...ivity If you want to exit the calibration press the ESC key the meter will immediately return to the measurement mode Temperature Calibration During the measurement process if the temperature reading...

Страница 16: ...Enter key the display shows the updated calibration information 4 Press the ESC key the meter returns to the measurement mode Conductivity TDS Salinity Resistivity Measurement 1 Press the Mode key in...

Страница 17: ...During the measurement process press the MI key to store the measured value the meter will show a reminder as follow Recalling from memory 1 1 Press the MR key in the measurement mode the meter shows...

Страница 18: ...1 In the measurement mode press and hold the key for 3 seconds to enter the setup menu 2 2 Press the or key to select the Clear Stored Data 2 3 Press the Enter key the cursor change to highlight 2 4...

Страница 19: ...communication between the meter and the computer has been established 5 Click the OK button to confirm 6 Click the Receive button the stored data automatically transfer to computer Interval recording...

Страница 20: ...84 S cm 1413 S cm 12 88mS cm 111 8mS cm TDS Range 0 00mg L 100 0g L Max 200g L Accuracy 1 F S TDS Factor 0 01 1 00 Default 0 5 Salinity Range 0 00 80 00ppt 0 00 42 00psu 0 00 8 00 Accuracy 1 F S Measu...

Страница 21: ...TA Conductivity at Temperature A CTB Conductivity at Temperature B TA Temperature A TB Temperature B 1 Press and hold the oC key for 3 seconds to enter the temperature setting 2 Press the or key to se...

Страница 22: ...to speed up these efforts Warranty The warranty period for meter is one year from the date of shipment Above warranty does not cover the sensor and calibration solutions Out of warranty products will...

Отзывы: