background image

Hazardous Substance Statement 

Bante  Instruments  is  committed  to  the  reduction  and  eventual  elimination  of  all  hazardous  substances  in  both  the  manufacturing  process  and  finished 
products  we  supply.  We  have  an  active  manufacturing  and  procurement  program  to  minimize  and  eliminate  the  use  of  harmful  heavy  metals  such  as 
cadmium, lead, mercury and the like. New technologies and design parameters are also promoting these efforts and we expect to have little or no such 
materials in our product in the coming years. We welcome our customer suggestions on how to speed up these efforts. 
 

 

 
 

Warranty 

The warranty period for meter is one year from the date of shipment. Above warranty does not cover the sensor and calibration solutions. Out of warranty 
products will be repaired on a charged basis. The warranty on your meter shall not apply to defects resulting from: 

 

Improper or inadequate maintenance by customer. 

 

Unauthorized modification or misuse. 

 

Operation outside of the environment specifications of the products. 

For more information, please contact the nearest authorized distributor. 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

 
BANTE INSTRUMENTS CO., LTD 
E-mail: [email protected] 
Phone: +8621-6404-1598 
Website: www.bante-china.com 
 

The information in this document is subject to change without notice. Copyright ©  Bante Instruments, 2018. All rights reserved. 

Содержание A151

Страница 1: ...A151 Benchtop Conductivity Meter Instruction Manual BANTE INSTRUMENTS CO LTD...

Страница 2: ...ent meets following conditions Relative humidity is less than 80 Ambient temperature is greater than 0 C and less than 60 C No potential electromagnetic interference The following list describes the s...

Страница 3: ...seconds Locks the measured value Resume measuring Sends data to a printer or computer Stores current reading to memory Increase value or scroll up through the menu item Views the calibration report o...

Страница 4: ...e arm automatically rises or falls you need to adjust the screw until arm locate at any position 1 Remove the plastic cover from the electrode arm 2 Use the screwdriver to tighten the screw moderately...

Страница 5: ...and dry Connecting the Temperature Probe 1 Place the temperature probe into the circular hole of the electrode arm 2 Insert the phone plug to the connector socket labeled C Ensure the connector is fu...

Страница 6: ...temperature compensation type and coefficient Linear 2 1 C Non linear Pure Water Coefficient Enable When the option is enabled pure water coefficient will be applied automatically for ultra pure water...

Страница 7: ...the default parameter The meter provides two methods for parameter settings Press the or key to modify the value press the Enter key to confirm If the cursor appears below the first digit press the or...

Страница 8: ...n is necessary Note non linear correction can only be performed at temperature range from 0 C to 36 C If the temperature value is out of above range the meter will show a warning To calculate the temp...

Страница 9: ...re is used to lock a measurement endpoint If enabled the meter will automatically sense a stable reading and lock the measurements The HOLD icon appears on the display Press the Meas key the meter res...

Страница 10: ...y and wait for entering the correct digits Press the or key to input the password press the Enter key to confirm Once you have successfully entered the setup menu selecting the Disable The password wi...

Страница 11: ...the meter is now switched to the automatic temperature compensation mode Manual Temperature Compensation If the meter does not detect a temperature probe the MTC icon will show on the display indicat...

Страница 12: ...lect a standard value close to the sample value you are measuring The meter will automatically detect these conductivity standard solutions and prompt the user to calibrate the meter When the calibrat...

Страница 13: ...ically shows current calibration standard e g 1413 S cm 1 5 If necessary press the or key to modify the calibration value 1 6 Press the Enter key the Calibrating icon shows on the display 1 7 Wait for...

Страница 14: ...ed the display will show Calibration Point 2 The meter prompts you to continue with second point calibration 2 3 Rinse the conductivity electrode with distilled water Place the electrode into the next...

Страница 15: ...ivity If you want to exit the calibration press the ESC key the meter will immediately return to the measurement mode Temperature Calibration During the measurement process if the temperature reading...

Страница 16: ...Enter key the display shows the updated calibration information 4 Press the ESC key the meter returns to the measurement mode Conductivity TDS Salinity Resistivity Measurement 1 Press the Mode key in...

Страница 17: ...During the measurement process press the MI key to store the measured value the meter will show a reminder as follow Recalling from memory 1 1 Press the MR key in the measurement mode the meter shows...

Страница 18: ...1 In the measurement mode press and hold the key for 3 seconds to enter the setup menu 2 2 Press the or key to select the Clear Stored Data 2 3 Press the Enter key the cursor change to highlight 2 4...

Страница 19: ...communication between the meter and the computer has been established 5 Click the OK button to confirm 6 Click the Receive button the stored data automatically transfer to computer Interval recording...

Страница 20: ...84 S cm 1413 S cm 12 88mS cm 111 8mS cm TDS Range 0 00mg L 100 0g L Max 200g L Accuracy 1 F S TDS Factor 0 01 1 00 Default 0 5 Salinity Range 0 00 80 00ppt 0 00 42 00psu 0 00 8 00 Accuracy 1 F S Measu...

Страница 21: ...TA Conductivity at Temperature A CTB Conductivity at Temperature B TA Temperature A TB Temperature B 1 Press and hold the oC key for 3 seconds to enter the temperature setting 2 Press the or key to se...

Страница 22: ...to speed up these efforts Warranty The warranty period for meter is one year from the date of shipment Above warranty does not cover the sensor and calibration solutions Out of warranty products will...

Отзывы: