Bante900P Portable Multiparameter Water Quality Meter
52
Addendum 2: Optional Ion Selective Electrodes
The meter is capable of connecting a variety of composite ion selective electrodes. The following list shows the selectable sensors and its range.
MODEL
ION
RANGE
F-US
Fluoride
1× 10
-6
M~ Saturation
CL-US
Chloride
5× 10
-5
~1M
Br-US
Bromide
5× 10
-7
~1M
Cn-US
Cyanide
8× 10
-6
~0.01M
Na-US
Sodium
4× 10
-6
~1M
NO3-US
Nitrate
7× 10
-6
~1M
Ca-US
Calcium
5× 10
-7
~1M
NH4-US
Ammonium
5× 10
-6
~0.1M
Cd-US
Cadmium
1× 10
-6
~0.1M
NH3-US
Ammonia
1× 10
-6
~1M
Cu-US
Cupric
1× 10
-6
~0.1M
I-US
Iodide
5× 10
-8
~1M
Pb-US
Lead
1× 10
-6
~1M
K-US
Potassium
7× 10
-6
~1M
Ag-US
Silver
1× 10
-7
~1M
S-US
Sulphide
1× 10
-7
~1M
Addendum 3: How to prepare the electrode storage solution
1. Dissolve 22.365 grams of potassium chloride reagent (KCL) in the 100mL distilled water.
2. Pour the 50mL of KCL solution into the beaker.
3. Pour the 50mL of pH4.01 standard buffer solution into the same beaker.
4. Stir the solution until the solution has thoroughly mixed. Preparation is completed.
Addendum 4: How to prepare the conductivity standard solutions
Place AR potassium chloride reagent in a 50mL beaker and dry in an oven for 3 to 5 hours at 105°C, then cool to room temperature in desiccator.
CALIBRATION SOLUTIONS
REAGENT
84µS/cm
Accurately weigh out 42.35mg of KCL and dissolve in 1000mL de-ionised water
1413µS/cm
Accurately weigh out 745.9mg of KCL and dissolve in 1000mL de-ionised water
12.88mS/cm
Accurately weigh out 7.45g of KCL and dissolve in 1000mL de-ionised water
111.8mS/cm
Accurately weigh out 74.5g of KCL and dissolve in 1000mL de-ionised water
Stir the solution until the reagent has thoroughly mixed.