Bante900P Portable Multiparameter Water Quality Meter
16
MULTI-POINT CALIBRATION:
2.1 Make sure that you have selected 2 to 5 points calibration in the setup menu.
2.2 Repeat steps 1.2 to 1.5 above. When the first calibration point is completed, the display will show “CAL2”. The meter prompts you to continue with
second point calibration.
2.3 Rinse the pH electrode with distilled water. Dip the electrode into the next standard buffer solution (e.g., pH4.01). The meter automatically senses
the current calibration solution and begins calibrating. “Calibration” indicator begins flashing.
2.4 Wait for the measured value to stabilize, the display shows “CAL3”. The meter prompts you to continue with third point calibration.
2.5 Repeat the steps 2.3 until the display shows “END”. The meter returns to measurement mode automatically. Calibration is completed.