Bante900P Portable Multiparameter Water Quality Meter
49
Conductivity
Range
0~20.00, 200.0, 2000µS/cm, 20.00, 200.0mS/cm
Accuracy
± 0.5% F.S
Calibration Points
1 to 5 points
Calibration Solutions
10µS/cm, 84µS/cm,1413µS/cm,12.88mS/cm,111.8mS/cm
Temperature Compensation
0~100°C, 32~212°F, Manual or Automatic
Temperature Coefficient
0.0~10.0%/°C
Compensation Modes
Linear or Pure Water
Cell Constant
K=0.1, 1, 10
Normalization Temperature
20°C or 25°C
TDS
Range
0~100ppt (Max. 200ppt, depending on factor setting)
Accuracy
± 1% F.S
TDS Factor
0.1~1.0 (Default 0.5)
Salinity
Range
0~10ppt (Max. 80ppt)
Accuracy
± 1% F.S
Measurement Modes
Seawater or Practical Salinity
Resistivity
Range
0~100MΩ
Accuracy
± 1% F.S
Resolution
0.01, 0.1, 1
Dissolved Oxygen
Range
0.00~20.00mg/L (or ppm)
Accuracy
± 0.2mg/L
Resolution
0.01mg/L
Calibration Points
1 or 2 points
Temperature Compensation
0~50°C, 32~122°F
Pressure Correction
60.0~112.5kPa, 450~850mmHg
Salinity Correction
0~50g/L
% Saturation of Oxygen
Range
0.0~200.0%
Accuracy
± 2.0%
Resolution
0.1%
Temperature
Range
0~105°C, 32~221°F
Accuracy
± 0.5°C, ± 0.9°F
Resolution
0.1°C
Calibration Points
1 point, Measured value ± 10°C