Bante900P Portable Multiparameter Water Quality Meter
48
Troubleshooting
LCD DISPLAY
CAUSE
CORRECTIVE ACTION
Electrode dried out
Soak the pH electrode in 3M KCL solution for 15 minutes
Soak the Ion Selective Electrode in the diluted standard solution
Soak the conductivity electrode in tap water for a few minutes
Measured value is out of range
Check the electrode whether clogged, dirty or broken
Incorrect calibration solutions
Using the fresh calibration solutions for calibration
Setting value does not match calibration solution
Reset the calibration value
DO electrolyte solution is depleted
Refilling electrolyte solution
Zero oxygen solution is contaminated
Replace the calibration solution
pH electrode is out of service life
Replace the pH electrode
Keypad is not working properly
Replace the batteries
Specifications
pH
Model
Bante900P
Range
-2.000~20.000pH
Accuracy
± 0.002pH
Resolution
0.1, 0.01, 0.001pH
Calibration Points
1 to 5 points
Calibration Solutions
USA, NIST, DIN, User-Defined Buffer
Temperature Compensation
0~100°C, 32~212°F, Manual or Automatic
mV
Range
-1999.9~1999.9mV
Accuracy
± 0.2mV
Resolution
0.1, 1mV
Calibration Points
1 point (Only for Relative mV mode)
Calibration Range
± 200mV
Ion
Range
0.001~19999ppm, mg/L, mol/L (Depending on range of ISE)
Accuracy
± 0.5% F.S (Monovalent), ± 1% F.S (Divalent)
Resolution
0.001, 0.01, 0.1, 1
Calibration Points
2 to 5 points
Calibration Solutions
0.001, 0.01, 0.1, 1, 10, 100, 1000, 10000ppm, mol/L, mg/L