Bante900P Portable Multiparameter Water Quality Meter
28
MULTI-POINT CALIBRATION:
2.1 Make sure that you have selected 2 to 5 points calibration in setup menu.
2.2 Repeat steps 1.2 to 1.5 above, when the first calibration point is completed, the display will show “CAL2”. The meter prompts you to continue with
second point calibration.
2.3 Rinse the conductivity electrode with distilled water. Dip the electrode into the next calibration solution, the meter automatically shows current
calibration standard (e.g., 12.88mS/cm).
2.4 Press ENTER key to confirm, “Calibration” indicator begins flashing.
2.5 Wait for the measured value to stabilize, the display shows “CAL3”. The meter prompts you to continue with third point calibration.
2.6 Repeat steps 2.3 to 2.4 above until the display shows “END”. The meter returns to measurement mode automatically. Calibration is completed.
MANUAL CALIBRATION:
Bante900P portable meter provides a quick manual calibration mode that allows user to easily calibrate the meter.
3.1 Record the cell constant value on the electrode (e.g., K=1.08).
Enter