Bante900P Portable Multiparameter Water Quality Meter
22
Ion Concentration Calibration
Bante900P portable multiparameter water quality meter
supports ion concentration calibration up to 5 points with minimum of 2 points, available calibration
points include the following options.
MEASUREMENT UNITS
CALIBRATION POINTS
ppm
0.001, 0.01, 0.1, 1, 10, 100, 1000, 10000
mg/L
0.001, 0.01, 0.1, 1, 10, 100, 1000, 10000
mol/L
0.001, 0.01, 0.1, 1, 10
mmol/L
0.001, 0.01, 0.1
To obtain accurate measurement results, we recommend that you perform ion calibration and measurement at same temperature. If you are not calibrate
the meter or calibration is not successfully, the display will always show "0.000".
1. Press MODE key until the meter shows
indicator.
2. If necessary, select the concentration unit and ion valence (Monovalent or Divalent) in the setup menu (Refer to Setup Menu section).
NOTE: The meter is capable of using the mg/L, ppm or mol/L as measurement unit of concentration. The factory default is ppm. If you had converted
concentration unit, the display will show “CAL” and flashing uninterruptedly. Press CAL key, the meter enters the calibration mode.