3.3 Keypad features
Main display
Laser indicator
Power indicator
Information
indicator
Fault indicator
Menu
Keypad lock
indicator
Cancel
Enter
Up, left, right,
down arrows
Figure 34
The
Aurora H2O
has seven keys: a menu key, four arrow keys,
a cancel
key, and an enter
key.
• Use the
menu
key to open the main menu on the display.
• Use the
arrow
keys to navigate among menu choices
and to increment/decrement numeric entries.
• Use the
cancel
key to cancel a numeric entry change,
or exit a menu.
• Use the
enter
key to accept a numeric entry or select
a menu option.
3.3.1 Indicator lights
If the
fault indicator
is lit, an instrument fault is detected.
A message will be displayed in the main display, top/right.
If the
information indicator
is lit, the instrument is still
operating, but a message will appear in the main display
top/right, with information about the instrument.
The
keypad lock indicator
will be lit if either: A) the keypad
lock-out switch, internal to the instrument, has been
engaged, or B) the instrument keypad has not been used
for a period of several minutes, engaging a software feature
to lock-out inadvertent key usage. Type (B) keypad lock-out
is overcome by pressing cancel, enter, cancel in sequence.
If the
laser indicator
is lit, the laser is powered and operating
normally. This indicator will be off if there is a laser-specific
fault. This indicator will also be off for a brief period when
the instrument is first powered. After initial power-up, this
indicator may blink several times as the laser temperature
is stabilized. The laser indicator will be lit constantly in
normal operation.
The
power indicator
is normally lit when the instrument
is powered on.
3.3.2 The magnetic stylus
Each of the keys can be selected using a hand-held
magnet called a magnetic stylus (see Figure 35 below),
which is included with the meter. By touching the clear
window at a key location, that key will be selected and will
flash a red light to verify the contact.
Figure 35: Magnetic stylus
3.3.3 The default display
Figure 36 below shows the default display of the
Aurora H2O
window.
Primary
Alt 2
Alt 1
Figure 36: Default display
25
Содержание Panametrics Aurora H2O
Страница 2: ...ii...
Страница 4: ...no content intended for this page proceed to next page iv...
Страница 20: ...no content intended for this page 10...
Страница 32: ...no content intended for this page 22...
Страница 65: ...2 AuroraView boots up and displays a screen similar to Figure 55 below Figure 55 AuroraView main screen 55...
Страница 71: ...Figure 74 Exporting a simplified image Figure 75 Selecting enhanced metafile 61...