2. Click alarms
config
This window enables the user to configure the alarm status within the
AuroraView
application. This feature allows you to
remotely configure
Aurora H2O
alarms, which are used only with modbus RTU digital output. The
AuroraView
alarms are
shown in Figure 58 and Figure 59 below and Figure 60 on page 57.
Figure 58: Alarms configuration
Toggle alarms
on/off
Set alarm trigger drop-down
Alarm type:
set point,
out of band,
in band
Limits
Alarm “read”
indication
(see below)
Datalog
alarm
condition
on/off
Figure 59: Other measurements options
Figure 60: Alarm status indicators
3. Click outputs
config
(see Figure 61 and Figure 62 below).
Figure 61: Outputs configuration
Set output parameter
Set output type: 0-20mA or 4-20mA
Set zero or lower range;
set span or upper range
Figure 62: Other output options
4. Click scan
This section enables you to pick the
desired type of scan. The default scan
is the SPECTRA scan, which shows the
2f spectral scan. This is the processed
signal waveform that the
Aurora
H2O
uses to determine the moisture
concentration. Viewing this scan may
assist in troubleshooting. A typical
2f spectra scan is shown in Figure
64 below. Select the scan interval in
minutes to set the refresh rate at which
AuroraView
updates the scan plot. To
enter a scan interval, click on the click
to save scans periodically button, and
the screen shown in Figure 63 below
appears. Enter the interval and click on
continue to save or cancel to reject the
change.
Figure 63: Save scans periodically
Figure 64: Scan plot tab
57
Содержание Panametrics Aurora H2O
Страница 2: ...ii...
Страница 4: ...no content intended for this page proceed to next page iv...
Страница 20: ...no content intended for this page 10...
Страница 32: ...no content intended for this page 22...
Страница 65: ...2 AuroraView boots up and displays a screen similar to Figure 55 below Figure 55 AuroraView main screen 55...
Страница 71: ...Figure 74 Exporting a simplified image Figure 75 Selecting enhanced metafile 61...
Страница 80: ...no content intended for this page 70...