Figure 7: Unpacking the Aurora H2O
Chapter 2. Installation
2.1 Introduction
The
Aurora H2O
analyzer provides direct indication of
moisture concentration in natural gas. Temperature
and pressure sensors are used to provide high precision
enhancement. It may be installed in a wide variety of
environmental conditions, and meets the requirements for
operation in hazardous areas.
Be sure that the ambient temperature is at least 10°C
higher than the maximum dew/frost point temperature
you expect to measure. This will ensure that you will not
have liquid condensation in the sample transport line or
the
Aurora H2O
sampling system components. Heat tracing
the sample line will aid in elevating the sample temperature
above the dew point. An optional heater installed within
the enclosure is also available.
2.2 Bill of materials
The following items should have been received with
your shipment:
•
Aurora H2O
Unit
•
Aurora H2O
User’s manual on CD ROM
•
AuroraView
software on CD ROM
•
Aurora H2O
calibration data sheet
• Maintenance/accessories kit
• Optional: Pipeline insertion filter/regulator assembly
2.3 Unpacking
The
Aurora H2O
is shipped in a packing box with a
plywood base (see Figure 7 below), and the
Aurora H2O
analyzer is secured to the plywood base with mounting
bolts. Transport the shipping package with the plywood
base on the bottom, and according to the warning labels
on the exterior packaging. Open the packing box from the
top, and remove the foam packing material. Collect any
separate components such as the
AuroraView
CD, the user
manual CD, and other items contained in the shipment.
Remove the mounting studs at the bottom of the enclosure.
Using two people, lift the
Aurora H2O
from the bottom of
the enclosure and from the top of the enclosure where the
analyzer electronics is located. A typical
Aurora H2O
weighs
approximately 100 pounds (45 kg). Use a proper lifting
technique to avoid injury.
Check all the received components and record the model
numbers and serial numbers for your records. If anything is
missing, contact Panametrics immediately.
11
Содержание Panametrics Aurora H2O
Страница 2: ...ii...
Страница 4: ...no content intended for this page proceed to next page iv...
Страница 20: ...no content intended for this page 10...
Страница 32: ...no content intended for this page 22...
Страница 65: ...2 AuroraView boots up and displays a screen similar to Figure 55 below Figure 55 AuroraView main screen 55...
Страница 71: ...Figure 74 Exporting a simplified image Figure 75 Selecting enhanced metafile 61...
Страница 80: ...no content intended for this page 70...