Copyright 2013 Baker Hughes Company.
English–DPI
620G Instruction Manual | 45
Measure or Simulate a Resistance Temperature Detector (RTD)
c. Sine
•
AMPLITUDE - Select the Peak-to-Peak value.
•
OFFSET - Set an offset value (only applies when Sine and Triangle waveforms are
selected).
4.9 Measure or Simulate a Resistance Temperature Detector (RTD)
14 show CH1 set up to measure an RTD. A 4
-
wire
configuration gives the best accuracy; a 2
-
wire configuration has the lowest accuracy.
Figure
4
-
12: PT100 RTD Measure CH1 4-Wire (Range
-
200 to 850°C)
Figure
4
-
13: PT100 RTD Measure CH1 3-Wire (Range
-
200 to 850°C)
Содержание Druck DPI 620 Genii
Страница 1: ...Druck com DPI 620 Genii Advanced Modular Calibrator Instruction Manual...
Страница 2: ......
Страница 12: ...Copyright 2013 Baker Hughes Company x DPI 620G Instruction Manual English...
Страница 20: ...Copyright 2013 Baker Hughes Company 8 DPI 620G Instruction Manual English Chapter 1 Introduction...
Страница 38: ...Copyright 2013 Baker Hughes Company 26 DPI 620G Instruction Manual English Chapter 2 Instrument Operation...
Страница 62: ...Copyright 2013 Baker Hughes Company 50 DPI 620G Instruction Manual English Chapter 4 Electrical Tasks...
Страница 72: ...Copyright 2013 Baker Hughes Company 60 DPI 620G Instruction Manual English Chapter 5 Pressure Tasks...
Страница 84: ...Copyright 2013 Baker Hughes Company 72 DPI 620G Instruction Manual English Chapter 7 Data Logging...
Страница 110: ...Copyright 2013 Baker Hughes Company 98 DPI 620G Instruction Manual English Chapter 9 HART Operations...
Страница 126: ...Copyright 2013 Baker Hughes Company 114 DPI 620G Instruction Manual English Chapter 10 FOUNDATION Fieldbus...
Страница 162: ...Copyright 2013 Baker Hughes Company 150 DPI 620G Instruction Manual English Chapter 12 Calibration Procedures...
Страница 164: ...Copyright 2013 Baker Hughes Company 152 DPI 620G Instruction Manual English Chapter 13 General Specification...