Copyright 2013 Baker Hughes Company.
English–DPI
620G Instruction Manual | 149
Procedures: RTD
-
INTERFACE
Figure
12
-
11: Calibration – RTD
-
INTERFACE
5. Check the calibration is correct:
a. Select the applicable RTD
-
Interface measure function via the Calibrator Task Menu.
b. Apply the following values:
%FS: 0, 25, 50, 75, 100
c. Check the error is within limits. Refer to Table
19.
Table
12
-
19: RTD Resistance (Measure) Error Limits
Applied Resistance (Ω)
Calibrator Uncertainty (Ω)
Permitted DPI
620 Genii Error
(Ω)
0
0.0020
0.020
100
0.0020
0.032
200
0.0029
0.044
300
0.0041
0.056
400
0.0052
0.068
Содержание Druck DPI 620 Genii
Страница 1: ...Druck com DPI 620 Genii Advanced Modular Calibrator Instruction Manual...
Страница 2: ......
Страница 12: ...Copyright 2013 Baker Hughes Company x DPI 620G Instruction Manual English...
Страница 20: ...Copyright 2013 Baker Hughes Company 8 DPI 620G Instruction Manual English Chapter 1 Introduction...
Страница 38: ...Copyright 2013 Baker Hughes Company 26 DPI 620G Instruction Manual English Chapter 2 Instrument Operation...
Страница 62: ...Copyright 2013 Baker Hughes Company 50 DPI 620G Instruction Manual English Chapter 4 Electrical Tasks...
Страница 72: ...Copyright 2013 Baker Hughes Company 60 DPI 620G Instruction Manual English Chapter 5 Pressure Tasks...
Страница 84: ...Copyright 2013 Baker Hughes Company 72 DPI 620G Instruction Manual English Chapter 7 Data Logging...
Страница 110: ...Copyright 2013 Baker Hughes Company 98 DPI 620G Instruction Manual English Chapter 9 HART Operations...
Страница 126: ...Copyright 2013 Baker Hughes Company 114 DPI 620G Instruction Manual English Chapter 10 FOUNDATION Fieldbus...
Страница 162: ...Copyright 2013 Baker Hughes Company 150 DPI 620G Instruction Manual English Chapter 12 Calibration Procedures...
Страница 164: ...Copyright 2013 Baker Hughes Company 152 DPI 620G Instruction Manual English Chapter 13 General Specification...
Страница 165: ......